YS-828 Particulate Matter Protection Effect Tester

产品描述:

YS-828 Particulate Matter Protection Effect Tester【Scope of application】It is us

  • 产品型号

  • 发布时间

    2026-03-10
  • 企业性质

  • 浏览次数

7×24在线咨询

专业客服在线解答,为您解决遇到的问题!

立即咨询

详细介绍

YS-828 Particulate Matter Protection Effect Tester

【Scope of application】

It is used to quickly and accurately detect the protective effect of various flat masks and 3D masks on particulate matter.

【Relevant Standards】

  • GB/T 32610-2016 Technical specifications for daily protective masks 6.14 Protective effect Appendix B Test methods for particulate matter protection effect, GB/T 38880-2020 Technical specifications for children's masks and other standards.

【Scope of application】
It is used to quickly and accurately detect the protective effect of various flat masks and 3D masks on particulate matter.
【Compliance with standards】
GB/T 32610-2016 Technical specification for daily protective masks 6.14 Protective effect Appendix B Test method for particulate matter protection effect,
GB/T 38880-2020 Technical specifications for children's masks and other standards.
InstrumentFeatures
1. The photometer sensor is used to detect the concentration of upstream and downstream particles to ensure the accuracy, stability, speed and effectiveness of sampling.
2. The system test flow rate is mainly provided by the outside to provide dry and clean compressed air. There is a voltage stabilization device installed inside to ensure the stability of the detection flow, and the automatic control system is simple, fast and stable.
3. Adopt Collision multi-nozzle design, which can test salt and oily particulate matter on the same instrument to ensure fast and stable adjustment of mist concentration.
4. It can be equipped with GB∕T 38880-2020 children's mask test sports silicone head mold.
5. Configure artificial respiration simulator, which can set the expiratory volume and respiratory rate arbitrarily.
Technical parameters
1. The size of the protective effect test chamber is about (length×width× height): 600mm× 600mm×560mm;
2. Test medium NaCl particulate matter;
2.1. The initial concentration of the effective space of the test chamber: (20~30) mg/m3, and the change during the test process is not more than 10%;
2.2. Aerodynamic particle size distribution of particulate matter: (0.02~2) μm;
2.3. Median mass diameter: 0.6μm;
3. Test medium: DEHS;
3.1. The initial concentration of the effective space of the test chamber: :(20~30) mg/m3, and the change during the test process is not more than 10%;
3.2. Aerodynamic particle size distribution of particulate matter: (0.02~2) μm;
3.3. Median mass diameter: 0.3μm;
4. a photometer sensor;
4.1. Gas collection flow rate: (1~2) L/min;
4.2. Detection frequency: 60 times/min;
4.3. Measurement range and accuracy of particulate matter concentration: (0.001~100) mg/m3, accuracy 1%;
4.4. The inhalation gas sampling tube is in the nostril;
4.5. The position of the ambient air sampling tube is 2cm away from the mouth and nose of the mask;
5. Matching head mold: one standard head mold (medium);
6. Breathing simulator:
6.1. sinusoidal airflow;
6.2. Respiratory rate: (5~40) times/min;
6.3. Respiratory flow: (5~80±1) L/min;
7. Power requirements: 220V, 50Hz, 1KW;
8. Air source requirements: 0.5MPa, 120L/Min;
9. Overall dimensions (length mm× width mm× height mm): 820×840×1260;
10. Weight: 62Kg for the main unit / 34Kg for the protective effect test module;